FDA Recall Terminated

IntelliVue MultiMeasurement Server (MMS)/Multimeasurement Server, physiological patient monitoring system. Model: M3001A

Recall: Z-1487-06 · Initiated July 24, 2006

Recall

Recall Number
Z-1487-06
Event Number
35785
Firm
Philips Medical Systems
FEI Number
1218950
Product Code
DQA
Status
Terminated
Root Cause
Other
Initiated
July 24, 2006
Posted
September 12, 2006
Terminated
August 11, 2020
Address
3000 Minuteman Road, Andover, MA, 01810-1032

Description

IntelliVue MultiMeasurement Server (MMS)/Multimeasurement Server, physiological patient monitoring system. Model: M3001A

Reason

Patient monitor may display inaccurate reading when the Disposable Sp02 Sensor is not attached

Action

Philips is using the contract services of Sterling, Indianapolis, IN to contact all consignees regarding the correction/upgrade to the device. The letter notification dated 7/21/06 was mailed UPS Delivery Tracking the week of 07/24/06. Additionally, Philips included with the Urgent Device Correction Notice a procedure to mitigate risk until the upgrade is installed.

Distribution

Worldwide, including USA, Canada, Argentina, Australia, Austria, Belgium, Brazil, Czech, Croatia, China, Denmark, Finland, France, Germany, Israel, Italy, Jordan, Kuwait, Luxemburg, Netherlands, Oman, Norway, Poland, Portugual, Poland, Russia, South Africa, Spain, Sweden, Swizerland, Turkey, UAE, Yemen, UK, Zimbabwe, Zambia, India, Indonesia, Japan, Korea, Philipines, Malaysia, Singapore, Sri Lanka, Taiwan, Vietnam, Venezuela.

Quantity

48,699 units