IntelliVue MultiMeasurement Server (MMS)/Multimeasurement Server, physiological patient monitoring system. Model: M3001A
Recall
- Recall Number
- Z-1487-06
- Event Number
- 35785
- Firm
- Philips Medical Systems
- FEI Number
- 1218950
- Product Code
- DQA
- Status
- Terminated
- Root Cause
- Other
- Initiated
- July 24, 2006
- Posted
- September 12, 2006
- Terminated
- August 11, 2020
- Address
- 3000 Minuteman Road, Andover, MA, 01810-1032
Description
IntelliVue MultiMeasurement Server (MMS)/Multimeasurement Server, physiological patient monitoring system. Model: M3001A
Patient monitor may display inaccurate reading when the Disposable Sp02 Sensor is not attached
Philips is using the contract services of Sterling, Indianapolis, IN to contact all consignees regarding the correction/upgrade to the device. The letter notification dated 7/21/06 was mailed UPS Delivery Tracking the week of 07/24/06. Additionally, Philips included with the Urgent Device Correction Notice a procedure to mitigate risk until the upgrade is installed.
Worldwide, including USA, Canada, Argentina, Australia, Austria, Belgium, Brazil, Czech, Croatia, China, Denmark, Finland, France, Germany, Israel, Italy, Jordan, Kuwait, Luxemburg, Netherlands, Oman, Norway, Poland, Portugual, Poland, Russia, South Africa, Spain, Sweden, Swizerland, Turkey, UAE, Yemen, UK, Zimbabwe, Zambia, India, Indonesia, Japan, Korea, Philipines, Malaysia, Singapore, Sri Lanka, Taiwan, Vietnam, Venezuela.
48,699 units