FDA Recall
Terminated
Alveolus Tracheobronchial Stent Technology System (TB-STS), Manufactured for: Alveolus Inc., 401 North Tryon Street, Charlotte, North Carolina USA 28202
Recall: Z-1487-04
·
Initiated April 22, 2004
Recall
- Recall Number
- Z-1487-04
- Event Number
- 29886
- Firm
- Alveolus, Inc.
- FEI Number
- 3004039296
- Product Code
- FWN
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 22, 2004
- Posted
- September 29, 2004
- Terminated
- January 12, 2005
- Address
- 401 N Tryon St, Fl 10th, Charlotte, NC, 28202-2108
Description
Alveolus Tracheobronchial Stent Technology System (TB-STS), Manufactured for: Alveolus Inc., 401 North Tryon Street, Charlotte, North Carolina USA 28202
Reason
The tracheal stent may collapse resulting in blockage of the air passage because the stents were not properly coated with the polyurethane material to the very end of the stent.
Action
Consignees were notified by telephone on 4/22/2004.
Distribution
FL, SC, TX