Altair 330 lbs w/o Hanger Bar, REF: 55200H-NA, 55220H-NA, and Roomer S 220 kg w/o Hanger Bar, REF: 55300H-NA, patient lifts
Recall
- Recall Number
- Z-1486-2024
- Event Number
- 94140
- Firm
- Human Care Hc Sweden Ab Arstaangsvagen 21 B Stockholm Sweden
- FEI Number
- 3006291016
- Product Code
- FSA
- Status
- Open, Classified
- Root Cause
- Process control
- Initiated
- February 23, 2024
- Posted
- April 5, 2024
Description
Altair 330 lbs w/o Hanger Bar, REF: 55200H-NA, 55220H-NA, and Roomer S 220 kg w/o Hanger Bar, REF: 55300H-NA, patient lifts
Patient lift hanger bar may come loose, which could potentially pose risks to patients.
On 2/23/24, Field Safety Notices were emailed to customers who were asked to do the following: 1) Checking the fixation by carefully pulling on the shaft. Should you observe any deviation or irregularity, please inform us promptly. 2) Read and understand this letter and provide a confirmation via email to the address [email protected] In case of recognizing any deviation or irregularity during the fixation control or if in need of additional information email [email protected] On 5/6/24, Field Safety Notices were emailed to customers with additional instructions for distributors and dealers to distribute the notice to all customers who have purchased the impacted products.
US Nationwide distribution in the states of TX, IL, OK, CA.