ACCU-CHEK FlexLink Plus, subcutaneous infusion set. Part Number 05511003001. AC FLEXLINK PLUS SET 8/40 10PK. An infusion set for the subcutaneous infusion of insulin administered with micro dosage insulin pumps. Made in Germany Manufactured for and Distributed by Roche Insulin Delivery Systems Inc. Fishers, IN 46037, USA Toll Free: 1-800-280-7801 Roche Diabetes Care AG 3401 Burgdorf, Switzerland
Recall
- Recall Number
- Z-1486-2011
- Event Number
- 58015
- Firm
- Roche Insulin Delivery Systems Inc.
- FEI Number
- 1000116163
- Product Code
- FRN
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- February 24, 2011
- Posted
- March 10, 2011
- Terminated
- July 27, 2011
- Address
- 11800 Exit 5 Pkwy, Ste 120, Fishers, IN, 46037-7989
Description
ACCU-CHEK FlexLink Plus, subcutaneous infusion set. Part Number 05511003001. AC FLEXLINK PLUS SET 8/40 10PK. An infusion set for the subcutaneous infusion of insulin administered with micro dosage insulin pumps. Made in Germany Manufactured for and Distributed by Roche Insulin Delivery Systems Inc. Fishers, IN 46037, USA Toll Free: 1-800-280-7801 Roche Diabetes Care AG 3401 Burgdorf, Switzerland
There is the potential to kink or bend the cannula when inserting the ACCU-CHEK FlexLink Plus infusion set. This can lead to under delivery and elevation of blood glucose levels.
Roche Insulin Delivery Systems sent urgent Product Recall letters, dated 2/21/2011. The letter identified the affected product and described the issue regarding kinked or bent cannulas. The firm requests that customers discontinue use of the ACCU-CHEK FlexLink Plus infusion set and return the unused stock. Customers are to contact their health care provider to determine if changes to their pump therapy are needed and how to temporarily continue insulin pump therapy without the ACCU-CHEK FlexLink Plus infusion set. The attached Customer Reply Card should be completed and return via mail or fax per the instructions on the card. The letter also lists some additional information pertaining to signs and symptoms. Questions regarding this recall should be directed to ACCU-CHEK Customer Care at 1-800-688-4578, 24 hours a day, 7 days a week.
Worldwide Distribution -- US and Canada.
1,454,918 million worldwide