FDA Recall Terminated

Boston Scientific Fort Guide Wire, Floppy Marker Wire, 300cm x 2 cm, REF 34949-02, for use in intravascular interventional procedures.

Recall: Z-1486-2009 · Initiated May 12, 2009

Recall

Recall Number
Z-1486-2009
Event Number
52098
Firm
Boston Scientific Corporation
FEI Number
3002095335
Product Code
DQX
Status
Terminated
Root Cause
Process control
Initiated
May 12, 2009
Posted
June 24, 2009
Terminated
December 11, 2011
Address
1 Scimed Pl, Maple Grove, MN, 55311-1565

Description

Boston Scientific Fort Guide Wire, Floppy Marker Wire, 300cm x 2 cm, REF 34949-02, for use in intravascular interventional procedures.

Reason

Product packaged incorrectly. One batch/lot of wires, that should contain radiopaque marker bands, is missing the marker bands; and a second batch/lot of wires, which should not contain radiopaque marker bands, has the marker bands. Since the marker bands issue would not be visually apparent to the physician there is a potential for prolongation or delay of the procedure in order to exchange the

Action

Consignees were sent a "Boston Scientific Urgent Medical Device Recall" letter dated may 19, 2009. The letter was addressed to "Dear Risk Manger/ Field Action Contact". The letter described the problem, customer steps for recall and listing of product batches / lots. The letter requested consignees to return the Reply Verification Tracking Form.

Distribution

USA distribution: AR, MO, CA, FL, and NV.

Quantity

Total of 5 boxes (5 units /box) ,