FDA Recall Open, Classified

Atellica CH Urine Albumin (UAlb). Material Number: 11537225

Recall: Z-1484-2026 · Initiated January 20, 2026

Recall

Recall Number
Z-1484-2026
Event Number
98370
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2432235
Product Code
JIR
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
January 20, 2026
Posted
March 2, 2026
Address
511 Benedict Ave, Tarrytown, NY, 10591-5005

Description

Atellica CH Urine Albumin (UAlb). Material Number: 11537225

Reason

Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.

Action

Siemens Healthineers notified consignees on about 01/20/2026 via mailed FedEx letter titled Urgent Medical Device Correction. Consignees were informed that as temporary mitigation, for any sample with a UAlb result above 5.0 mg/dL (50 mg/L), reanalyze the sample with a x10 dilution. If the diluted result is consistent with the initial result, either result may be reported. If the diluted result is inconsistent with the initial result, the diluted result should be reported. If you prefer to automate this dilution, please refer to your internal laboratory procedures. Consignees were instructed to review the notification with their Medical Director to determine the appropriate course of action, including evaluation of any previously generated results, if applicable, and complete and return the provided Field Correction Effectiveness Check Form. A follow-up communication will be provided when Customer Actions are no longer required.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, Curacao,¿St¿Eus, Czech¿Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Korea, Latvia, Malaysia, Netherlands, New Zealand, Norway, Portugal, Romania, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, United¿Kingdom.

Quantity

4,885 units