FDA Recall Open, Classified

RayStation Software Version Numbers version numbers 8B, 9A, 9B, 10A, 10B, 11A, 11B, 12A, 12B, 2023B, 2024A including some service packs.

Recall: Z-1482-2024 · Initiated February 23, 2024

Recall

Recall Number
Z-1482-2024
Event Number
94153
Firm
RAYSEARCH LABORATORIES AB Eugeniavagen 18c Stockholm Sweden
FEI Number
3007774465
Product Code
MUJ
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
February 23, 2024
Posted
April 5, 2024

Description

RayStation Software Version Numbers version numbers 8B, 9A, 9B, 10A, 10B, 11A, 11B, 12A, 12B, 2023B, 2024A including some service packs.

Reason

A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.

Action

On February 26, 2024, the firm notified customers of the issue via Urgent Field Safety Notice: Medical Device Correction letters via email. Actions to be taken by the user: 1. Educate planning staff and all users about the possibility of the described use error. 2. If the use error is deemed likely to occur at your clinic, take actions to ensure that proper review of assigned materials is performed before a patient model is used for treatment planning. 3. If you wish to have the Silicon chemical element removed from the material list, contact RaySearch support for assistance. This is especially recommended for clinics that perform ion beam treatments, where the harm caused by the use error could be significant. 4. Inspect your product and identify all installed units with the above software version number(s). 5. Confirm you have read and understood this notice by replying to the notification email. The Silicon chemical element will not be included in the next version of RayStation/RayPlan, scheduled for market release in April 2024 (subject to market clearance in some markets). If requested by the customer, it can be removed from the material list of existing installations. This notice needs to be passed on to all those who need to be aware within your organization. Maintain awareness of this notice as long as any affected version is in use. Thank you for your cooperation, and we apologize for any inconvenience. For regulatory information, please contact [email protected].

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Canada, Colombia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, France, Germany, Hong Kong S.A.R., Hungary, India, Indonesia, Israel, Italy, Japan, Korea, Malaysia, Morocco, Netherlands, New Zealand, Norway, Poland, P.R. China, Qatar, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, United Kingdom.

Quantity

154 units