Fabius GS Premium, anesthesia machine, catalog no. 8607000 Product Usage: The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2).
Recall
- Recall Number
- Z-1482-2018
- Event Number
- 79773
- Firm
- Draeger Medical, Inc.
- FEI Number
- 2517967
- Product Code
- BSZ
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- March 1, 2018
- Posted
- May 8, 2018
- Terminated
- April 16, 2019
- Address
- 3135 Quarry Rd, Telford, PA, 18969-1042
Description
Fabius GS Premium, anesthesia machine, catalog no. 8607000 Product Usage: The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2).
The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.
On March 1, 2018, the recalling firm notified affected customers via an Urgent Medical Device Recall letter. The recalling firm stated that customers would receive replacement motors for potentially affected devices. Until the replacement occurred, customers were instructed to continue to operate their devices with the usual attention and make sure that a manual resuscitator for emergency ventilation is kept ready according to the instructions for use. Customers were also supplied with a reply card to indicate that they had received the notification. Customers with questions about the letter were directed to call Michael Kelhart between 8AM-4:30PM EST at 1-800-437-2437 (press 1 at the prompt, then 32349). Questions regarding the operation and/or servicing of the Draeger Fabius anesthesia machine were directed to DraegerService Technical Support between the hours of 8:00 AM 8:00 PM EST at 1-800-437-2437 (press 4 at the prompt then 4 again).
US Nationwide Distribution to accounts in 21 states: AZ, CA, CO, ID, IL, KS, MA, MI, MS, NC, NY, OH, OK, PA, RI, TN, TX, UT, VA, WA, WI; and PR.
27