FDA Recall Terminated

MaxGuard Extension Set, Model No. ME2064, MP9009C, MP9209, MP9254C The MaxGuard extension sets are used to administer fluids from a container to a patient s vascular system through a needle or catheter inserted into the patient s artery or vein. The extension sets may incorporate components that aid in the prevention of accidental needle sticks. The set s components are commonly found on intravascular administration sets and extension sets

Recall: Z-1482-2017 · Initiated February 20, 2017

Recall

Recall Number
Z-1482-2017
Event Number
76440
Firm
CareFusion 303, Inc.
FEI Number
2016493
Product Code
FPA
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
February 20, 2017
Posted
February 28, 2017
Terminated
October 29, 2018
Address
10020 Pacific Mesa Blvd, San Diego, CA, 92121-4386

Description

MaxGuard Extension Set, Model No. ME2064, MP9009C, MP9209, MP9254C The MaxGuard extension sets are used to administer fluids from a container to a patient s vascular system through a needle or catheter inserted into the patient s artery or vein. The extension sets may incorporate components that aid in the prevention of accidental needle sticks. The set s components are commonly found on intravascular administration sets and extension sets

Reason

CareFusion is recalling the MaxGuard Extension Set due to reports of leaks occurring with the 0.2 micron filter.

Action

CareFusion sent an Urgent Medical Device Recall Notification letter dated February 2017 to all affected customers inform them that CareFusion is recalling the MaxGuard Extension Set model codes ME20164, MP9009-C, MP9209, MP9254-C with multiple lot numbers. CareFusion has received reports of leaks occurring with the 0.2 micron filter. Customers were informed that leakages on an extension can cause delay of infusion, interruption of infusion, exposure to medication or hazardous infusates, underinfusion or air in line. Customers were instructed to discontinue use of the affected product and contact CareFusion for a replacement lot. The letter informs the customers of the actions to be taken and for recall related questions to contact CareFusion Support Center at (888) 562-6018. Customers with adverse event reports are instructed to contact Customer Advocacy at (888) 812-3266. Customers were also instructed to immediately complete and return to CareFusion the enclosed, pre-addressed and postage paid, Recall Response Card.

Distribution

Nationwide Distribution

Quantity

12,850 units