Hill-Rom brand VersaCare Bed; Model P3200.

Recall: Z-1482-04 · Initiated August 24, 2004

Recall

Recall Number
Z-1482-04
Event Number
29897
Firm
Hill-Rom, Inc.
FEI Number
1824206
Product Code
FNL
Status
Terminated
Root Cause
Other
Initiated
August 24, 2004
Posted
September 21, 2004
Terminated
January 31, 2005
Address
125 E Pearl St, Batesville, IN, 47006

Description

Hill-Rom brand VersaCare Bed; Model P3200.

Reason

Pivot bolts on the sleep deck may become loose and fall out, allowing the head section to shift or to drop to one side, and a grounding warning label is missing.

Action

Recall letters dated August 24, 2004 were sent to consignees advising them to inspect and monitor the beds until Hill-Rom can visit and make corrections.

Distribution

Nationwide, Brazil, Canada, Hong Kong and United Arab Emirates.

Quantity

5048