FDA Recall Open, Classified

LUMINOS Lotus Max (VF11)

Recall: Z-1481-2023 · Initiated November 23, 2022

Recall

Recall Number
Z-1481-2023
Event Number
92229
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
OWB
Status
Open, Classified
Root Cause
Radiation Control for Health and Safety Act
Initiated
November 23, 2022
Posted
May 23, 2023
Address
40 Liberty Blvd, Malvern, PA, 19355-1418

Description

LUMINOS Lotus Max (VF11)

Reason

Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC 60601-2-43. There is no impact on workflow or diagnosis. The error can occur only in fluoroscopy systems with a second (overhead) X-ray tube.

Action

Siemens Medical Solutions USA, Inc., will provide software updates to the affected systems with software versions VF10 and VF11 with software via Update Instructions (UI) XP007/23/S, XP004/23/S, XP003/23/S, and XP001/23/S

Distribution

US Nationwide - Worldwide Distribution

Quantity

704 units in the United States (1945 units worldwide)