FDA Recall Terminated

Fabius MRI, anesthesia machine, catalog no. 8607300 Product Usage: The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2). This device can also be used in MRI scanner rooms with magnets 1.5 tesla and 3 tesla at a fringe field strength of 40 mtesla or less.

Recall: Z-1481-2018 · Initiated March 1, 2018

Recall

Recall Number
Z-1481-2018
Event Number
79773
Firm
Draeger Medical, Inc.
FEI Number
2517967
Product Code
BSZ
Status
Terminated
Root Cause
Process control
Initiated
March 1, 2018
Posted
May 8, 2018
Terminated
April 16, 2019
Address
3135 Quarry Rd, Telford, PA, 18969-1042

Description

Fabius MRI, anesthesia machine, catalog no. 8607300 Product Usage: The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2). This device can also be used in MRI scanner rooms with magnets 1.5 tesla and 3 tesla at a fringe field strength of 40 mtesla or less.

Reason

The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.

Action

On March 1, 2018, the recalling firm notified affected customers via an Urgent Medical Device Recall letter. The recalling firm stated that customers would receive replacement motors for potentially affected devices. Until the replacement occurred, customers were instructed to continue to operate their devices with the usual attention and make sure that a manual resuscitator for emergency ventilation is kept ready according to the instructions for use. Customers were also supplied with a reply card to indicate that they had received the notification. Customers with questions about the letter were directed to call Michael Kelhart between 8AM-4:30PM EST at 1-800-437-2437 (press 1 at the prompt, then 32349). Questions regarding the operation and/or servicing of the Draeger Fabius anesthesia machine were directed to DraegerService Technical Support between the hours of 8:00 AM 8:00 PM EST at 1-800-437-2437 (press 4 at the prompt then 4 again).

Distribution

US Nationwide Distribution to accounts in 21 states: AZ, CA, CO, ID, IL, KS, MA, MI, MS, NC, NY, OH, OK, PA, RI, TN, TX, UT, VA, WA, WI; and PR.

Quantity

8