FDA Recall Terminated

ACCU-CHEK FlexLink Plus, subcutaneous infusion set. Part Number 05511054001. AC FLEXLINK PLUS 8MM CANNULA. An infusion set for the subcutaneous infusion of insulin administered with micro dosage insulin pumps. Made in Germany Manufactured for and Distributed by Roche Insulin Delivery Systems Inc. Fishers, IN 46037, USA Toll Free: 1-800-280-7801 Roche Diabetes Care AG 3401 Burgdorf, Switzerland

Recall: Z-1481-2011 · Initiated February 24, 2011

Recall

Recall Number
Z-1481-2011
Event Number
58015
Firm
Roche Insulin Delivery Systems Inc.
FEI Number
1000116163
Product Code
FRN
Status
Terminated
Root Cause
Device Design
Initiated
February 24, 2011
Posted
March 10, 2011
Terminated
July 27, 2011
Address
11800 Exit 5 Pkwy, Ste 120, Fishers, IN, 46037-7989

Description

ACCU-CHEK FlexLink Plus, subcutaneous infusion set. Part Number 05511054001. AC FLEXLINK PLUS 8MM CANNULA. An infusion set for the subcutaneous infusion of insulin administered with micro dosage insulin pumps. Made in Germany Manufactured for and Distributed by Roche Insulin Delivery Systems Inc. Fishers, IN 46037, USA Toll Free: 1-800-280-7801 Roche Diabetes Care AG 3401 Burgdorf, Switzerland

Reason

There is the potential to kink or bend the cannula when inserting the ACCU-CHEK FlexLink Plus infusion set. This can lead to under delivery and elevation of blood glucose levels.

Action

Roche Insulin Delivery Systems sent urgent Product Recall letters, dated 2/21/2011. The letter identified the affected product and described the issue regarding kinked or bent cannulas. The firm requests that customers discontinue use of the ACCU-CHEK FlexLink Plus infusion set and return the unused stock. Customers are to contact their health care provider to determine if changes to their pump therapy are needed and how to temporarily continue insulin pump therapy without the ACCU-CHEK FlexLink Plus infusion set. The attached Customer Reply Card should be completed and return via mail or fax per the instructions on the card. The letter also lists some additional information pertaining to signs and symptoms. Questions regarding this recall should be directed to ACCU-CHEK Customer Care at 1-800-688-4578, 24 hours a day, 7 days a week.

Distribution

Worldwide Distribution -- US and Canada.

Quantity

1,454,918 million worldwide