Regius Model 370 Digital Radiography Konica Minolta
Recall: Z-1481-06
·
Initiated July 14, 2006
Recall
- Recall Number
- Z-1481-06
- Event Number
- 36109
- Firm
- Konica Minolta Medical Imaging USA, Inc.
- FEI Number
- 2241281
- Product Code
- MQB
- Status
- Terminated
- Root Cause
- Other
- Initiated
- July 14, 2006
- Posted
- August 31, 2006
- Terminated
- September 5, 2006
- Address
- 411 Newark Pompton Tpke, Wayne, NJ, 07470-6657
Description
Regius Model 370 Digital Radiography Konica Minolta
Reason
A hand grip column, even though locked in position, moved downward about 3 cm while a patient was holding the hand grip for lateral exposure. When the hand grip moved down, the patient who was holding it lost his or her balance and almost fell. This problem only occurred in Japan with one system.
Action
Konica Minolta USA contacted the site of the sole US user by phone on 7/14/06, followed by e-mail and certified mail/return receipt.
Distribution
Texas
Quantity
1 in US