FDA Recall Terminated

Regius Model 370 Digital Radiography Konica Minolta

Recall: Z-1481-06 · Initiated July 14, 2006

Recall

Recall Number
Z-1481-06
Event Number
36109
Firm
Konica Minolta Medical Imaging USA, Inc.
FEI Number
2241281
Product Code
MQB
Status
Terminated
Root Cause
Other
Initiated
July 14, 2006
Posted
August 31, 2006
Terminated
September 5, 2006
Address
411 Newark Pompton Tpke, Wayne, NJ, 07470-6657

Description

Regius Model 370 Digital Radiography Konica Minolta

Reason

A hand grip column, even though locked in position, moved downward about 3 cm while a patient was holding the hand grip for lateral exposure. When the hand grip moved down, the patient who was holding it lost his or her balance and almost fell. This problem only occurred in Japan with one system.

Action

Konica Minolta USA contacted the site of the sole US user by phone on 7/14/06, followed by e-mail and certified mail/return receipt.

Distribution

Texas

Quantity

1 in US