FDA Recall Open, Classified

Revogene, Catalog no. 610210. IVD test instrument

Recall: Z-1479-2022 · Initiated June 22, 2022

Recall

Recall Number
Z-1479-2022
Event Number
90549
Firm
Meridian Bioscience Inc
FEI Number
1524213
Product Code
OOI
Status
Open, Classified
Root Cause
Device Design
Initiated
June 22, 2022
Address
3471 River Hills Dr, Cincinnati, OH, 45244-3023

Description

Revogene, Catalog no. 610210. IVD test instrument

Reason

When a run is aborted, the cooling period protection does not occur and the user is able to open the lid prior to the completion of the cooling period, exposing the user to heated instrument components. Incidental contact with heated instrument components may result in an injury such as minor burns.

Action

On June 22, 2022, the firm notified customers via email. Customers were instructed that if a run is aborted and an error code presented, they SHOULD NOT open the instrument. DO NOT touch the retention ring or the inside of the lid as they may be hot. The firm recommends that customers wait approximately 10 minutes to allow the instrument to cool prior to opening the lid and removing the test devices. The firm has developed a firmware solution to prevent the lid from being opened after an aborted run prior to the completion of the cooling period. The Meridian Bioscience Technical Services team will contact you to set up an appointment for firmware installation once available.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Chile, Italy, Qatar, and Switzerland.

Quantity

459 (394 US, 65 OUS)