Medtronic DxTerity Diagnostic Catheter, JL 4.0, REF DXT5JL40, marked with 6F color code on box.
Recall
- Recall Number
- Z-1479-2018
- Event Number
- 79591
- Firm
- Medtronic Vascular
- FEI Number
- 1220452
- Product Code
- DQO
- Status
- Terminated
- Root Cause
- Packaging
- Initiated
- March 13, 2018
- Posted
- April 20, 2018
- Terminated
- May 7, 2021
- Address
- 35-37A Cherry Hill Drive, Danvers, MA, 01923-2565
Description
Medtronic DxTerity Diagnostic Catheter, JL 4.0, REF DXT5JL40, marked with 6F color code on box.
The color coded French size indicator on the outer box may incorrectly reflect a 6 French size, rather than the correct 5 French size. The catheters inside the carton are the correct DXT5JL40 item, the inner pouch has correct 5F labeling, information and configuration details.
Medtronic initiated consignee communications on 13-Mar- 2018 to Risk Managers and Health Care Professionals. US notifications were sent via 2-Day UPS delivery. OUS communications were sent per locally approved methods. Consignees are asked to quarantine and return any unused potentially affected product to Medtronic.
US and AUSTRIA, Canada, Germany, ROMANIA, SLOVENIA,
855 units