FDA Recall
Terminated
Allen Medical C-Flex, Polar Head Positioner Catalog Number: A-70700-A1 Product is a medical device used to maintain the position of an anesthetized patient during surgery.
Recall: Z-1478-2008
·
Initiated November 27, 2007
Recall
- Recall Number
- Z-1478-2008
- Event Number
- 46781
- Firm
- Allen Medical Systems, Inc. One Post Office Square
- FEI Number
- 1221538
- Product Code
- CCX
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- November 27, 2007
- Posted
- August 20, 2008
- Terminated
- August 26, 2008
- Address
- Acton, MA, 01720
Description
Allen Medical C-Flex, Polar Head Positioner Catalog Number: A-70700-A1 Product is a medical device used to maintain the position of an anesthetized patient during surgery.
Reason
Handle jamming in the open position, preventing the device from adequately supporting the head during spinal surgery. Possible delay in surgery, or injury.
Action
Allen Medical notified accounts by telephone on 11/08/07 and the Sales Representative. The Hospitals confirmed that the device would be sent back for evaluation and upgraded.
Distribution
MA, MS
Quantity
2