FDA Recall Terminated

ADVIA Centaur CA 125 II Assay packaged in 100 and 500 test kits as follows: Product code: 09427226 (128532) 100 test kit Product code: 01678114 (128533) 500 Test kit

Recall: Z-1477-2011 · Initiated December 14, 2010

Recall

Recall Number
Z-1477-2011
Event Number
57427
Firm
Siemens Healthcare Diagnostics, Inc
FEI Number
1219913
Product Code
LTK
Status
Terminated
Root Cause
Process control
Initiated
December 14, 2010
Posted
March 2, 2011
Terminated
April 12, 2013
Address
333 Coney St, East Walpole, MA, 02032-1516

Description

ADVIA Centaur CA 125 II Assay packaged in 100 and 500 test kits as follows: Product code: 09427226 (128532) 100 test kit Product code: 01678114 (128533) 500 Test kit

Reason

Positive bias (40-55%) with patient results greater than the upper limit of normal and less than the upper limit of normal

Action

Siemens Healthcare Diagnostics on 12 /14/10 issued Urgent Device Recall notice via Federal Express to accounts. This notice informs the customer of the recall and the actions that need to be taken. The Urgent Device Recall notice directs customers to immediately discontinue the use of ADVIA Centaur CA 125 II assay kits that contain CA 125 II reagent lot 144. It also recommends that patient results obtained with this reagent lot be reviewed. A confirmation fax-back form is included to ensure that all customers worldwide have been notified and have complied with the mandatory action. Technical questions, please contact the Technical Solutions Center at 1-877-229-3711

Distribution

Worldwide distribution, including USA, Canada, Argentina, Austria, Belgium, China, Denmark, France, Germany, Great Britain, Greece, India, Italy, Japan, Korea, Lavitia, Lithuania, Malyasia, Mexico, Netherlands, Norway, Peru, Philippines, Poland, Portugal, Romania, Saudi Arabia, Switzerland, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Czech Republic, Taiwan, Turkey, Uganda, and Venezuela

Quantity

2960 units