FDA Recall
Open, Classified
Campy CVA Medium 100/PK, R01272
Recall: Z-1474-2026
·
Initiated February 5, 2026
Recall
- Recall Number
- Z-1474-2026
- Event Number
- 98403
- Firm
- Remel, Inc
- FEI Number
- 1924669
- Product Code
- JSJ
- Status
- Open, Classified
- Root Cause
- Under Investigation by firm
- Initiated
- February 5, 2026
- Posted
- February 26, 2026
- Address
- 12076 Santa Fe Trail Dr, Lenexa, KS, 66215-3519
Description
Campy CVA Medium 100/PK, R01272
Reason
Customer complaints report low to no recovery of Campylobacter Jejuni ATCC 33291 on identified lot of medium.
Action
On February 5, 2026, the firm notified affected customers via Urgent Medical Device Recall letter. Customers were instructed to review test results and destroy any remaining inventory of the affected lot.
Distribution
US Nationwide distribution in the states of AZ, CO, IL, NJ, OH, TN, WA.
Quantity
97