SafeLight Fiber Optic Cable: Clear, 5MM X 10FT 3.05M, REF: 0233-050-100; Gray, 5MM X 10FT 3.05M, REF: 0233-050-200
Recall
- Recall Number
- Z-1474-2023
- Event Number
- 91891
- Firm
- Stryker Corporation
- FEI Number
- 2936485
- Product Code
- FCW
- Status
- Open, Classified
- Root Cause
- Employee error
- Initiated
- March 2, 2023
- Posted
- April 28, 2023
- Address
- 5900 Optical Ct, San Jose, CA, 95138-1400
Description
SafeLight Fiber Optic Cable: Clear, 5MM X 10FT 3.05M, REF: 0233-050-100; Gray, 5MM X 10FT 3.05M, REF: 0233-050-200
Fiber optic cables assembled with not enough epoxy on the proximal end, creating light output issues, which could cause a clinical delay or procedure being converted to open.
On 3/2/23 recall notices were mailed to customers who were asked to do the following: 1. Inform individuals within your organization who need to be aware of this action. 2. Identify, remove, and quarantine affected devices. 3. If affected product is found, call customer service at 1-800-624-4422 (Option 3) or email [email protected] to arrange for product return. 4. Complete and return the Business Reply Form Customers with questions can contact [email protected]
Worldwide - US Nationwide distribution including in the states of PA, MS, IL, OH, CA, TX, MO, IN, KS, OK, MT, MA, NE, AZ, NV, NJ, IA, WV, MD, KY, FL, NM, NY, CO, WA, NC, VA, LA, ND, WY, HI SC, UT, OR, SD, AR, TN, AK, CT, GA, ID, VT, NH, MN, WI, DE, RI, AL, MI and the countries of NL, PE, AT, CA, TW, PH, SG, KR, CN.