FDA Recall Terminated

SlingBar 360. Designed to meet the needs for lifting humans.

Recall: Z-1474-2013 · Initiated April 30, 2013

Recall

Recall Number
Z-1474-2013
Event Number
64550
Firm
Hill-Rom, Inc.
FEI Number
1824206
Product Code
FSA
Status
Terminated
Root Cause
Device Design
Initiated
April 30, 2013
Posted
June 5, 2013
Terminated
November 28, 2014
Address
1069 State Route 46 East, Batesville, IN, 47006-7520

Description

SlingBar 360. Designed to meet the needs for lifting humans.

Reason

Retrospective review found 4 sling bars that may present a potential for impalement if a patient, caregiver or bystander were to fall onto the hooks based on a simulated test method. In effort to eliminate any risk of injury, Hill-Rom is requesting products be removed from field use and replaced with an updated design.

Action

On 4/30/13 an URGENT MEDICAL DEVICE RECALL CORRECTION was issued to all consignees detailing the hazards and actions that are needed. Included with this notification is a Response Form/Receipt and a Replacement Guide, complete and return the form to Hill-Rom as soon as possible. Hill-Rom will send you the applicable number of sling bars to you at no cost. After you receive the new sling bars, we request that you discard the old ones.

Distribution

Worldwide Distribution-USA (nationwide) including the states of AZ, CA, CO, FL, IL, IN, KS, LA, MA, MI, MN, MO, NC, NH, NJ, NV, OH, OK, PA, SC, TN, UT, WA, and WI, and the countries of Canada, Australia, Austria, Belgium, Switzerland, Germany Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, Italy, Japan, Netherlands, Norway, New Zealand, Portugal, Sweden, and Turkey.

Quantity

16