VITROS CRP Slides: (1) Cat # 192 6740, 250 slides per box, and (2) Cat # 809 7990, 90 slides per box. Firm on the label: Ortho-Clinical Diagnostics, Inc., Rochester, NY 14626. The CRP Slides are for use on the Vitros 250/250AT and/or Vitros 950/950AT Clinical Chemistry Systems. For in vitro diagnostic use. Vitros CRP slides quantitatively measure C-reactive protein concentration in serum and plasma.
Recall
- Recall Number
- Z-1473-04
- Event Number
- 29887
- Firm
- Ortho-Clinical Diagnostics
- FEI Number
- 1000136573
- Product Code
- DCK
- Status
- Terminated
- Root Cause
- Other
- Initiated
- July 28, 2004
- Posted
- September 18, 2004
- Terminated
- June 9, 2005
- Address
- 100 Indigo Creek Dr, Rochester, NY, 14626-5101
Description
VITROS CRP Slides: (1) Cat # 192 6740, 250 slides per box, and (2) Cat # 809 7990, 90 slides per box. Firm on the label: Ortho-Clinical Diagnostics, Inc., Rochester, NY 14626. The CRP Slides are for use on the Vitros 250/250AT and/or Vitros 950/950AT Clinical Chemistry Systems. For in vitro diagnostic use. Vitros CRP slides quantitatively measure C-reactive protein concentration in serum and plasma.
The Immuno-rate (IR) wash detection algorithm may result in inappropriate wash detections.
Letters dated 7/28/2004. Distributors instructed to destroy product. Users instructed to use revised IR wash tolerance values (provided with letter). A Calibration Diskette, Data Release Version 5384 will also be provided to end users.
DOMESTIC distribution (nationwide): Product shipped to 6 distributors, 140 hospitals, and 6 U.S. government accounts. FOREIGN distribution: Product shipped to 22 Ortho-Clinical/J&J affiliates in Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Peru, Ecuador, India, Japan, Mexico, Puerto Rico, Panama, Singapore, Venezuela, England, France, Germany, Italy and Spain. The foreign affiliates further distributed product.
Cat 192 6740 - 5,912 cartons x 250 slides; Cat #809 7990 - 10,344 x 90 slides