Replacement Kit, Power PCBA, REF 17345-001. only distributed outside the US. Pulmonetic Systems, 17400 Medina Rd., Suite 100, Minneapolis, MN 55447, USA.
Recall
- Recall Number
- Z-1470-2008
- Event Number
- 46850
- Firm
- Cardinal Health 203, Inc
- FEI Number
- 3002617992
- Product Code
- CBK
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- February 22, 2008
- Posted
- August 13, 2008
- Terminated
- August 16, 2012
- Address
- 17400 Medina Rd, Ste 100, Minneapolis, MN, 55447-1341
Description
Replacement Kit, Power PCBA, REF 17345-001. only distributed outside the US. Pulmonetic Systems, 17400 Medina Rd., Suite 100, Minneapolis, MN 55447, USA.
The failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
An Urgent Medical Device Recall letter dated February 22, 2008 was sent to ventilator consignees/owners and distributors. The letter describes the issue, identifies the affected devices, states that the devices will be corrected by Cardinal Health and provides direction to verify ventilator operation and audible alarm function daily. Service Centers were provided information to identify affected printed circuit board assembly replacement kits' lot numbers and asked to return them. An "Update" was mailed 03/04/2008 to identify additional ventilator serial numbers that also need correction.
Worldwide Distribution --- USA including Puerto Rico and countries of Algeria, Argentina, Australia, Austria, Bahamas, Belgium, Brazil, Canada, Chile, China Peoples Republic, Colombia, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Israel, Italy, Japan, Malaysia, Mexico, New Zealand, Norway, Panama, Paraguay, Philippines, Portugal, Saudi Arabia, Serbia & Montenegro, Singapore, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam, and West Indies.
175 (OUS only)