FDA Recall
Terminated
OEC FlexiView 8800, Digital Mobile Imaging System, GE Medical Systems
Recall: Z-1469-2017
·
Initiated February 21, 2017
Recall
- Recall Number
- Z-1469-2017
- Event Number
- 76568
- Firm
- GE OEC Medical Systems, Inc
- FEI Number
- 1720753
- Product Code
- OXO
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- February 21, 2017
- Posted
- February 27, 2017
- Terminated
- April 24, 2017
- Address
- 384 N Wright Brothers Dr, Salt Lake City, UT, 84116-2862
Description
OEC FlexiView 8800, Digital Mobile Imaging System, GE Medical Systems
Reason
OEC FlexiView 8800 workstation power cable assembly issue which can result in a loss of imaging functionality due to intermittent power cord failures.
Action
On February 21, 2017, an Urgent Medical Device Correction letter was sent to one consignee with an affected unit; a field service representative hand delivered the customer communication.
Distribution
One consignee in CA.
Quantity
1