FDA Recall Terminated

Zimmer VERSYS Hip System Femoral Stem, press-fit, enhanced taper, 12/14 neck taper, collarless, size 15, 140 mm stem length, tivanium TI-6AL-4V alloy, sterile; Cat. no. 7861-15-04 (00786101504).

Recall: Z-1469-06 · Initiated August 2, 2006

Recall

Recall Number
Z-1469-06
Event Number
35943
Firm
Zimmer Inc.
FEI Number
1000220733
Product Code
LWJ
Status
Terminated
Root Cause
Other
Initiated
August 2, 2006
Posted
August 31, 2006
Terminated
May 4, 2007
Address
345 E Main St, Warsaw, IN, 46580-2746

Description

Zimmer VERSYS Hip System Femoral Stem, press-fit, enhanced taper, 12/14 neck taper, collarless, size 15, 140 mm stem length, tivanium TI-6AL-4V alloy, sterile; Cat. no. 7861-15-04 (00786101504).

Reason

Lack of assurance of proper metal fatigue strength due to a metal grain structure anomaly.

Action

Consignees were notified via letter dated 8/2/06 to cease use of and to return the products. Zimmer distributors were to examine stock and return product on hand.

Distribution

Worldwide, including USA, Australia, China, Denmark, Japan, Jordan, Latin Ameria, Russia, Singapore and Switzerland.

Quantity

181 of all devices