Convenience kits containing Tego Connectors: Medline DIALYSIS DRESSING CHANGE SKU EBSI1498 Medline DIALYSIS DRESSING CHANGE KIT SKU EBSI1107 EBSI1596 EBSI1781 EBSI1812 Centurion DIALYSIS/PHERESUS CVC DRSG CHANGE KIT SKU DT22630 Medline DIALYSIS/PHERESIS DRESSING CHANGE KIT SKU EBSI1470 Centurion VASCATH DRESSING CHANGE TRAY SKU DT22845
Recall
- Recall Number
- Z-1467-2026
- Event Number
- 98333
- FEI Number
- 1417592
- Product Code
- OXQ
- Status
- Open, Classified
- Root Cause
- Process control
- Initiated
- January 8, 2026
- Posted
- February 24, 2026
- Address
- 3 Lakes Dr, Medline Industries, LP, Northfield, IL, 60093-2753
Description
Convenience kits containing Tego Connectors: Medline DIALYSIS DRESSING CHANGE SKU EBSI1498 Medline DIALYSIS DRESSING CHANGE KIT SKU EBSI1107 EBSI1596 EBSI1781 EBSI1812 Centurion DIALYSIS/PHERESUS CVC DRSG CHANGE KIT SKU DT22630 Medline DIALYSIS/PHERESIS DRESSING CHANGE KIT SKU EBSI1470 Centurion VASCATH DRESSING CHANGE TRAY SKU DT22845
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
On January 8, 2026, the firm notified affected customers of the product issue through email and first class mail. Customers were instructed to request stickers to over-label affected convenience kits with a warning label indicating to remove the affected component and replace it with product from supply. No kits are to returned.
US Nationwide distribution.
31,848