FDA Recall Terminated

FP1000 Cell Preparation System Part Number 624922

Recall: Z-1466-06 · Initiated June 26, 2006

Recall

Recall Number
Z-1466-06
Event Number
35778
Firm
Beckman Coulter Inc
FEI Number
2050012
Product Code
LXG
Status
Terminated
Root Cause
Other
Initiated
June 26, 2006
Posted
August 31, 2006
Terminated
April 26, 2012
Address
200 S Kraemer Blvd, Brea, CA, 92822-6208

Description

FP1000 Cell Preparation System Part Number 624922

Reason

During the cleaning cycle performed during the shutdown procedure of the Beckman Coulter FP1000 Cell Preparation System, fluid (diluted bleach) may drip from the probe in the location of the back reagent rack potentially resulting in bleach and/or water dripping into the reagents contaminating them and resulting in possible incorrect results.

Action

A Product Corrective Action letter was sent on June 26, 2006 informing users that during the cleaning cycle performed during the shutdown procedure of the Beckman Coulter FP1000 Preparation System, fluid (diluted bleach) may drip from the probe in the location of the back reagent rack potentially resulting in bleach and/or water dripping into the reagents contaminating them and resulting in possible incorrect results. To reduce the risk of incorrect results from dripping fluid, users were advised to remove all patient samples, reagents, controls, and calibrators before initiating the shutdown cycle of the Beckman Coulter FP1000. Software will be modified in the next version to prevent the possibility of dripping bleach during the shutdown cycle. The Product Corrective Action letter is to be maintained as part of the customer''s laboratory Quality System documentation. The enclosed Response Form is to be completed and returned within 10 days to the recalling firm. Customers are also advised to contact the firm or their local Beckman Coulter Representative if there are questions.

Distribution

Nationwide and Canada

Quantity

11 units distributed