FDA Recall Terminated

Active Articulation E1. Intended for use with either primary or revision hip arthroplasty.

Recall: Z-1465-2013 · Initiated April 26, 2013

Recall

Recall Number
Z-1465-2013
Event Number
65209
Firm
Biomet, Inc.
FEI Number
1825034
Product Code
LPH
Status
Terminated
Root Cause
Labeling mix-ups
Initiated
April 26, 2013
Posted
June 5, 2013
Terminated
August 29, 2013
Address
56 E Bell Dr, Warsaw, IN, 46582-6989

Description

Active Articulation E1. Intended for use with either primary or revision hip arthroplasty.

Reason

Biomet is recalling Active Articulation E1 Hip Bearing implants. Biomet has initiated this action following an investigation which identified that outer label indicates two conflicting O.D. sizes while the inner labels indicate an incorrect O.D. size. The implant itself is etched with the correct O.D. size.

Action

Biomet sent an Urgent Medical Device Recall letter dated April 26, 2013 to consignees detailing that the outer label indicates two conflicting O.D. sizes while the inner labels indicate an incorrect O.D. size. This action requires the immediate location and discontinued use of the product and its return to Biomet. Ciustomers were advised to: Immediately locate and remove the identified device(s) listed from circulation . -Carefully follow the instructions on the enclosed "FAX Back Response Form" . - Fax a copy of the Response Form to 574-372-1683 prior to return of product. -Use priority carrier for your shipment. - If you have further distributed this product, you MUST notify hospital personnel of this action via the enclosed "Dear O.R. Manager" notice. This letter MUST be given to hospital personnel responsible for receiving recall notices. However, you are charged with the location and return of these products.

Distribution

Distributed in New York.

Quantity

8 units