FDA Recall Open, Classified

VITEK 2 - AST Test kits (in vitro diagnostic test cards for antimicrobial susceptibility): a) AST-N334, REF 418984; b) AST-GN95, REF 421982; c) AST-GN99, REF 423102; d) AST-N390, REF 423340; e) AST-N391, REF 423341; f) AST-N392, REF 423342; g) AST-N395, REF 423491; h) AST-N401, REF 423643; i) AST-N402, REF 423644; j) AST-N404, REF 423664; k) AST-N802, REF 423706; l) AST-N405, REF 423864; m) AST-N417, REF 423880; n) AST-N408, REF 423924; o) AST-N409, REF 423925; p) AST-N420, REF 424039; q) AST-N423, REF 424042; r) AST-N422, REF 424056; s) AST-N436, REF 424440; t) AST-N809, REF 424703; u) AST-N806, REF 424709; v) AST-N807, REF 424710; w) AST-N808, REF 424711; x) AST-N810, REF 424712; y) AST-N812, REF 424721

Recall: Z-1464-2024 · Initiated March 14, 2024

Recall

Recall Number
Z-1464-2024
Event Number
94264
Firm
Biomerieux Inc
FEI Number
1950204
Product Code
LON
Status
Open, Classified
Root Cause
Employee error
Initiated
March 14, 2024
Posted
April 25, 2024
Address
595 Anglum Rd, Hazelwood, MO, 63042-2320

Description

VITEK 2 - AST Test kits (in vitro diagnostic test cards for antimicrobial susceptibility): a) AST-N334, REF 418984; b) AST-GN95, REF 421982; c) AST-GN99, REF 423102; d) AST-N390, REF 423340; e) AST-N391, REF 423341; f) AST-N392, REF 423342; g) AST-N395, REF 423491; h) AST-N401, REF 423643; i) AST-N402, REF 423644; j) AST-N404, REF 423664; k) AST-N802, REF 423706; l) AST-N405, REF 423864; m) AST-N417, REF 423880; n) AST-N408, REF 423924; o) AST-N409, REF 423925; p) AST-N420, REF 424039; q) AST-N423, REF 424042; r) AST-N422, REF 424056; s) AST-N436, REF 424440; t) AST-N809, REF 424703; u) AST-N806, REF 424709; v) AST-N807, REF 424710; w) AST-N808, REF 424711; x) AST-N810, REF 424712; y) AST-N812, REF 424721

Reason

Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible results. The risk is present for isolates that have a result of MIC=0.5, 1, 2.

Action

BioMerieux issued an URGENT FIELD SAFETY NOTICE to its consignees on 03/14/2024 via email. The notice explained the issue, impact to patient, and provided the consignee (user) with alternate testing methods. BioMerieux issued another URGENT FIELD SAFETY NOTICE dated 07MAY2024 to its consignees on May 9, 2024, via email. This letter includes the statement, "however, no results for the Ceftriaxone (cro02n) should be reported outside the lab."

Distribution

Worldwide

Quantity

49215 units