FDA Recall Terminated

Ingenuity Core 128, Computed Tomography X-ray systems, Philips Medical Systems (Cleveland), Inc.

Recall: Z-1464-2014 · Initiated March 6, 2014

Recall

Recall Number
Z-1464-2014
Event Number
67763
Firm
Philips Medical Systems (Cleveland) Inc
FEI Number
1525965
Product Code
KPS
Status
Terminated
Root Cause
Software in the Use Environment
Initiated
March 6, 2014
Posted
April 16, 2014
Terminated
November 5, 2014
Address
595 Miner Rd, Cleveland, OH, 44143-2131

Description

Ingenuity Core 128, Computed Tomography X-ray systems, Philips Medical Systems (Cleveland), Inc.

Reason

A customer reported that after upgrading to software version 3.5.5 from 3.5.4, reconstructions would intermittently not complete.

Action

.The firm, Phillips, sent an " URGENT Medical Device Correction" letter dated 2013 March 06, to its costumers . The letter disc ribes the product, problem and actions to be taken. The costumers were instructed to Verify that all planned image reconstructions are completed, and Perform off-line reconstructions if expected images are missing. The firm will release A Field Change Order (FCO) to correct the issue with software version release 3.5.7.and A Philips Field Service Engineer will contact you to schedule the software installation at your site. If you need any further information or support concerning this issue, please contact your local Philips representative or local Philips Healthcare office. For North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377, Option 5: Enter Site ID or follow the prompts).

Distribution

Worldwide Distribution:US ( Nationwide) Including stages of MI, OR, PA, TX & WV; and Countries of: Australia, Bangladesh, Belgium, Brazil, Bulgaria, China, France, Germany, India, Indonesia, Ireland, Japan, Nicaragua, Oman, Palestine, Russia, South Korea, Spain, Switzerland, Thailand, Turkey & United Kingdom.

Quantity

116 units