FDA Recall
Terminated
nordicBrainEX, software versions 2.21 through 2.3.10
Recall: Z-1462-2022
·
Initiated June 16, 2022
Recall
- Recall Number
- Z-1462-2022
- Event Number
- 90480
- Firm
- NordicNeuroLab AS Mollendalsveien 65c Bergen Norway
- FEI Number
- 3006738448
- Product Code
- LLZ
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- June 16, 2022
- Terminated
- September 4, 2024
Description
nordicBrainEX, software versions 2.21 through 2.3.10
Reason
When exporting merged results from the BOLD, DSC, and DCE modules of nordicBrainEx, where both the underlay and overlay originated from a multiframe dataset, the resulting output may have left/right sides flipped.
Action
The firm sent a notice regarding a Critical Bug in nordicBrainEx version 2.2.1 to 2.3.10 on 06/16/2022 by email. The notice explained the problem and advised users to use extra caution when viewing the exported results. A software revision will be released to correct the issue.
Distribution
US Nationwide distribution.
Quantity
243 units