FDA Recall Terminated

AxSYM Drugs of Abuse/Toxicology Assay Disk Version 8.0; List Number: 3D54-08; Abbott La boratories, Diagnostics Division, Abbott Park, IL 60064

Recall: Z-1462-2008 · Initiated January 21, 2008

Recall

Recall Number
Z-1462-2008
Event Number
46740
Firm
Abbott Laboratories
FEI Number
1415939
Product Code
DKZ
Status
Terminated
Root Cause
Software change control
Initiated
January 21, 2008
Posted
May 14, 2008
Terminated
December 1, 2009
Address
100 Abbott Park Rd, Abbott Park, IL, 60064-3502

Description

AxSYM Drugs of Abuse/Toxicology Assay Disk Version 8.0; List Number: 3D54-08; Abbott La boratories, Diagnostics Division, Abbott Park, IL 60064

Reason

Error Message: If the Positive and Negative Interpretation Cutoff parameters (116/117) are edited after installation of the revised assay file, VRTX error #0002 in Task 40 is generated and the instrument locks up when it attempts to report an AxSYM Amphetamine/Methamphetamine II patient result.

Action

Consignees were sent a Product Correction-Immediate Action Required letter on 1/21/08. The letter provided users with the steps to take to temporarily correct the problem until the recalling firm is able to send a corrected assay file. Users were requested to return a customer reply letter acknowledging receipt of the letter. For additional information, contact 1-877-4ABBOTT.

Distribution

Worldwide Distribution including USA territory of Puerto Rico, and countries of Australia, Canada, Chile, Colombia, Germany, Mexico, New Zealand, South Korea, Taiwan, Thailand and Venezuela.

Quantity

1,270 disks