FDA Recall Open, Classified

Medline procedure kits labeled as: 1) BILLINGS CLINIC SC HEAD & NECK, Pack Number DYNJ33656C; 2) BLEPH/RHINO PACK, Pack Number DYNJ64669A; 3) DB ENT CSTM PACK-LF, Pack Number DYNJ33585C; 4) DENTAL PACK, Pack Number DYNJ54064; 5) EAR PACK, Pack Number DYNJ68678C; 6) ENT HEAD NECK PACK, Pack Number DYNJ49023D; 7) ENT PACK, Pack Number DYNJ42564B; 8) ENT PACK, Pack Number DYNJ50572; 9) ENT PACK, Pack Number DYNJ45175B; 10) ENT PACK, Pack Number DYNJ65585A; 11) EP PACK, Pack Number DYNJ59171; 12) FUNCTIONAL ENDO SINUS WEILER, Pack Number DYNJ58857A; 13) HEAD & NECK, Pack Number DYNJ902987F; 14) HEAD & NECK PACK, Pack Number DYNJ61295B; 15) HEAD & NECK PACK, Pack Number DYNJ30234B; 16) HEAD & NECK PACK, Pack Number DYNJ50383; 17) HEAD & NECK PACK, Pack Number DYNJ36634K; 18) HEAD & NECK PACK, Pack Number DYNJ66399; 19) HEAD AND NECK PACK-LF, Pack Number DYNJ38402I; 20) MAJOR ENT HARPER PACK-LF, Pack Number DYNJ0161710I; 21) NASAL SINUS PACK-LF, Pack Number DYNJ0751486N; 22) NE NASAL, Pack Number DYNJ03130D; 23) ORL PACK, Pack Number DYNJ63705B; 24) SINUS PACK, Pack Number DYNJ50110B; 25) SINUS PACK, Pack Number DYNJ50110B; 26) T & A PACK, Pack Number DYNJ58811A; 27) THORACIC PACK, Pack Number DYNJ30731M; 28) THYROIDECTOMIE DIAGNOSTIQUE-LF, Pack Number DYNJ42678A; 29) TONSILS & ADENOIDS, Pack Number DYNJ69465; 30) TRANSPHENOIDAL PACK, Pack Number DYNJ46556B;

Recall: Z-1461-2025 · Initiated February 11, 2025

Recall

Recall Number
Z-1461-2025
Event Number
96373
Firm
MEDLINE INDUSTRIES, LP Northfield
FEI Number
1417592
Product Code
OGR
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
February 11, 2025
Posted
March 26, 2025
Address
3 Lakes Dr, Northfield, IL, 60093-2753

Description

Medline procedure kits labeled as: 1) BILLINGS CLINIC SC HEAD & NECK, Pack Number DYNJ33656C; 2) BLEPH/RHINO PACK, Pack Number DYNJ64669A; 3) DB ENT CSTM PACK-LF, Pack Number DYNJ33585C; 4) DENTAL PACK, Pack Number DYNJ54064; 5) EAR PACK, Pack Number DYNJ68678C; 6) ENT HEAD NECK PACK, Pack Number DYNJ49023D; 7) ENT PACK, Pack Number DYNJ42564B; 8) ENT PACK, Pack Number DYNJ50572; 9) ENT PACK, Pack Number DYNJ45175B; 10) ENT PACK, Pack Number DYNJ65585A; 11) EP PACK, Pack Number DYNJ59171; 12) FUNCTIONAL ENDO SINUS WEILER, Pack Number DYNJ58857A; 13) HEAD & NECK, Pack Number DYNJ902987F; 14) HEAD & NECK PACK, Pack Number DYNJ61295B; 15) HEAD & NECK PACK, Pack Number DYNJ30234B; 16) HEAD & NECK PACK, Pack Number DYNJ50383; 17) HEAD & NECK PACK, Pack Number DYNJ36634K; 18) HEAD & NECK PACK, Pack Number DYNJ66399; 19) HEAD AND NECK PACK-LF, Pack Number DYNJ38402I; 20) MAJOR ENT HARPER PACK-LF, Pack Number DYNJ0161710I; 21) NASAL SINUS PACK-LF, Pack Number DYNJ0751486N; 22) NE NASAL, Pack Number DYNJ03130D; 23) ORL PACK, Pack Number DYNJ63705B; 24) SINUS PACK, Pack Number DYNJ50110B; 25) SINUS PACK, Pack Number DYNJ50110B; 26) T & A PACK, Pack Number DYNJ58811A; 27) THORACIC PACK, Pack Number DYNJ30731M; 28) THYROIDECTOMIE DIAGNOSTIQUE-LF, Pack Number DYNJ42678A; 29) TONSILS & ADENOIDS, Pack Number DYNJ69465; 30) TRANSPHENOIDAL PACK, Pack Number DYNJ46556B;

Reason

Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health.

Action

Medline Industries LP issued a recall notice to its consignees on 02/11/2025 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers.

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

Quantity

2161 kits