FDA Recall Terminated

Arkon Anesthesia Delivery System, Model 99999. Product Usage: Continuous Ventilator The Arkon Anesthesia Delivery System is designed to provide emergency oxygen, vaporized agent delivery and manual ventilation in the event of a power failure scenario. Clinicians are able to use Emergency O2 and manually ventilate the patient, providing gas and agent. However, the anesthesiologist will not have access to mechanical ventilation or ventilator monitoring.

Recall: Z-1460-2017 · Initiated February 8, 2017

Recall

Recall Number
Z-1460-2017
Event Number
76429
Firm
Del Mar Reynolds Medical, Ltd.
FEI Number
3002807992
Product Code
CBK
Status
Terminated
Root Cause
Software design
Initiated
February 8, 2017
Terminated
October 19, 2017
Address
2 Harforde Court John Tate Road, 1 Business Park Hertford United Kingdom

Description

Arkon Anesthesia Delivery System, Model 99999. Product Usage: Continuous Ventilator The Arkon Anesthesia Delivery System is designed to provide emergency oxygen, vaporized agent delivery and manual ventilation in the event of a power failure scenario. Clinicians are able to use Emergency O2 and manually ventilate the patient, providing gas and agent. However, the anesthesiologist will not have access to mechanical ventilation or ventilator monitoring.

Reason

Spacelabs has received reports of the Arkon Anesthesia Delivery System, Model 99999, shutting down without warning. The power supply has failed in a way that does not trigger the normal on battery alarms.

Action

On 08 February 2017, a customer letter will be sent via priority service, return receipt requested, to all U.S. customers. Records of notification will be maintained to ensure complete notification. On 15 February 2017, a customer letter (translated as necessary) will be emailed to all international subsidiaries and distributors of record. Records of notification will be maintained to ensure complete notification. Upgrade is expected to be available in April 2017; all affected Arkon Anesthesia Delivery Systems will receive a software upgrade onsite.

Distribution

Worldwide Distribution - US nationwide and internationally in the following countries: BOLIVIA, CANADA, CHINA, ECUADOR, FRANCE, IRELAND, MALAYSIA, MEXICO, MOROCCO, NETHERLANDS, PANAMA, PERU, POLAND, UNITED KINGDOM.

Quantity

325 devices sold to 35 US consignees and 76 devices sold to 14 international consignees