Acuity with Conebeam Computed Tomography; Radiation therapy simulation system; Model Number: H77.
Recall
- Recall Number
- Z-1460-2008
- Event Number
- 46919
- Firm
- Varian Medical Systems Oncology Systems
- FEI Number
- 2916710
- Product Code
- KPQ
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- January 3, 2008
- Posted
- July 30, 2008
- Terminated
- September 30, 2009
- Address
- 911 Hansen Way, Palo Alto, CA, 94304-1028
Description
Acuity with Conebeam Computed Tomography; Radiation therapy simulation system; Model Number: H77.
Incorrect Image Orientation: The image orientation tag may not be set correctly if the patient has been scanned with an orientation other than Head First Spine. When such images are imported to a treatment planning system, they may appear with either a mirrored or a rotated view.
Consignees were notified by an Urgent Medical Device Correction letter issued on 02/05/2008. The letter provided recommended actions for users to follow in order to avoid the software error. The letter informed users that a service pack is being created to correct the issue and that the recalling firm will notify users when it becomes available. For additional information, contact 888-827-4265.
Worldwide Distribution--USA including states of CA, MA and PA, and countries of UK, Chile, Belgium, Canada and Germany.
16 UNITS