FDA Recall Terminated

Acuity with Conebeam Computed Tomography; Radiation therapy simulation system; Model Number: H77.

Recall: Z-1460-2008 · Initiated January 3, 2008

Recall

Recall Number
Z-1460-2008
Event Number
46919
Firm
Varian Medical Systems Oncology Systems
FEI Number
2916710
Product Code
KPQ
Status
Terminated
Root Cause
Software design
Initiated
January 3, 2008
Posted
July 30, 2008
Terminated
September 30, 2009
Address
911 Hansen Way, Palo Alto, CA, 94304-1028

Description

Acuity with Conebeam Computed Tomography; Radiation therapy simulation system; Model Number: H77.

Reason

Incorrect Image Orientation: The image orientation tag may not be set correctly if the patient has been scanned with an orientation other than Head First Spine. When such images are imported to a treatment planning system, they may appear with either a mirrored or a rotated view.

Action

Consignees were notified by an Urgent Medical Device Correction letter issued on 02/05/2008. The letter provided recommended actions for users to follow in order to avoid the software error. The letter informed users that a service pack is being created to correct the issue and that the recalling firm will notify users when it becomes available. For additional information, contact 888-827-4265.

Distribution

Worldwide Distribution--USA including states of CA, MA and PA, and countries of UK, Chile, Belgium, Canada and Germany.

Quantity

16 UNITS