FDA Recall Terminated

S1730 Calibration Solution 2, 200 mL, packed in single unit plastic bottles, REF 944-025, used with ABL700 Series Blood Gas Analyzers.

Recall: Z-1460-06 · Initiated October 31, 2005

Recall

Recall Number
Z-1460-06
Event Number
35663
Firm
Radiometer America Inc
FEI Number
1523456
Product Code
CHL
Status
Terminated
Root Cause
Other
Initiated
October 31, 2005
Posted
August 29, 2006
Terminated
April 17, 2012
Address
810 Sharon Dr, Westlake, OH, 44145-1598

Description

S1730 Calibration Solution 2, 200 mL, packed in single unit plastic bottles, REF 944-025, used with ABL700 Series Blood Gas Analyzers.

Reason

Calibration solution for ABL700 Blood Gas Analyzers is labeled with an incorrect bar code. The bar code identified on the Cal Solution 2 product is actually the bar code for the rinse solution.

Action

Representatives of the recalling firm were informed of this problem and were asked to visit the firm''s customers and retrieve any of the suspect product found in inventory at each customer location for customer credit. The representatives were instructed to complete this action via a ''Management Cover e-mail issued on 10/31/2005. According to the Management Cover Memo, these actions were scheduled for completion by firm representatives on or before 12/15/2005.

Distribution

The recalled product was distributed to customers in following states nationwide: AL, AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI, WV and WY

Quantity

638/200 mL plastic bottles.