Covidien Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT
Recall
- Recall Number
- Z-1458-2024
- Event Number
- 94128
- FEI Number
- 1219930
- Product Code
- GDW
- Status
- Open, Classified
- Root Cause
- Under Investigation by firm
- Initiated
- February 14, 2024
- Posted
- April 1, 2024
- Address
- LP 60 Middletown Ave, Covidien, North Haven, CT, 06473-3908
Description
Covidien Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT
Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.
On February 14, 2024, the firm notified domestic customers and notified international customers starting February 19, 2024 of the issue via URGENT: MEDICAL DEVICE RECALL letters. Actions to be taken by the customer: 1. Identify and quarantine all unused impacted product listed above. 2. See Attachment A for guidance on identifying affected lots. 3. Return all quarantined devices to Medtronic. Please contact [email protected] for the Return Goods Authorization (RGA). 4. Credit for the returned affected devices will be issued based on the RGA number. 5. If purchased from a distributor, contact your distributor directly to arrange for the return of the devices back to your distributor. 6. Please complete and return the enclosed Customer Confirmation Form to [email protected] even if you do not have unused inventory. 7. Pass on this notice to all those who need to be aware within your organization or to any organization where the listed affected lots may have been transferred or distributed. Adverse events or quality problems experienced with this product should be reported to the FDA and Medtronic: Online at http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm, or call FDA at 1-800-FDA-1088 (1-800-332-1088) and Medtronic Technical Services at 800-255-6774, option 2 to provide information regarding those events. We regret any inconvenience this may cause. We are committed to patient safety and appreciate your prompt attention to this matter. If you have any questions regarding this communication, please contact your Medtronic Field Representative or Customer Service at 800-962-9888 Option 2.
Worldwide distribution - US Nationwide and the countries of France, Hong Kong, Israel and Japan.
3,023 units