FDA Recall Terminated

The Panorama HFO system has the Traceable Item Identification (12NC): 9896 030 17741-SW R 2.5.1 (shipped as R 2.5.1, upgraded to R 2.5.3 via FCO). 9896 030 19641-SW R 2.5.3 9896 030 19861-SW R 2.6.1 9896 030 19862-SW R 2.6.1 9896 030 19863-SW R 2.6.3

Recall: Z-1458-2011 · Initiated February 5, 2010

Recall

Recall Number
Z-1458-2011
Event Number
57561
Firm
Philips Medical Systems North America Co. Phillips
FEI Number
1000524572
Product Code
LNH
Status
Terminated
Root Cause
Other
Initiated
February 5, 2010
Posted
March 1, 2011
Terminated
August 10, 2011
Address
22100 Bothell Everett Hwy, Bothell, WA, 98021-8431

Description

The Panorama HFO system has the Traceable Item Identification (12NC): 9896 030 17741-SW R 2.5.1 (shipped as R 2.5.1, upgraded to R 2.5.3 via FCO). 9896 030 19641-SW R 2.5.3 9896 030 19861-SW R 2.6.1 9896 030 19862-SW R 2.6.1 9896 030 19863-SW R 2.6.3

Reason

In the ExamCard MobiFlex the Patient Position is "Prone" which must be "Supine". The result is mislabeling of left and right side of the patient. If this is not detected by the operator there is a risk of wrong treatment or the need of additional X-RAY during treatment.

Action

Philips began sendng out the URGENT-Device Correction Notice letter dated February 5, 2010, ro all affected consignees on February 10, 2010. The consignees were advised to change the patient position from "Prone" immediately to "Supine" and store the corrected ExamCard in the hospital database if they use the ExamCard, MobiFlex dual Injection. Consignees were instructed to call Philips Healthcare Call Center at 800-722-9377 #5 (then follow prompts) and also reference "FCO 78100318" with any communication.

Distribution

Nationwide Distribution including the following states: AL, CA, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, NE, NH, NJ, NY, OH, OR, SC, TN, TX, VA, VT, WI, and WV.AL, CA, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, NE, NH, NJ, NY, OH, OR, SC, TN, TX, VA, VT, WI, and WV. There was no foreign consignees.

Quantity

82 units