Freestyle Navigator Continuous Glucose Monitor System, Clinical System Kit, Packaging. Abbott Item Number: PRT 03415-010, distributed by Abbott Diabetes Care Inc, Alameda, CA. Medical device indicated for continually recording interstitial fluid glucose levels in people ages 18+ with diabetes mellitus for improving diabetes management. The device is indicated to be used in hospital and home settings to aid health care.
Recall
- Recall Number
- Z-1458-2010
- Event Number
- 52635
- Firm
- Abbott Diabetes Care, Inc.
- FEI Number
- 3002803444
- Product Code
- MDS
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- April 14, 2009
- Posted
- April 30, 2010
- Terminated
- March 31, 2011
- Address
- 1360 South Loop Road, Alameda, CA, 94502-7000
Description
Freestyle Navigator Continuous Glucose Monitor System, Clinical System Kit, Packaging. Abbott Item Number: PRT 03415-010, distributed by Abbott Diabetes Care Inc, Alameda, CA. Medical device indicated for continually recording interstitial fluid glucose levels in people ages 18+ with diabetes mellitus for improving diabetes management. The device is indicated to be used in hospital and home settings to aid health care.
Potential failure or inaccurate readings--the device's plastic housing near the battery compartment may crack, and allow moisture to enter.
Notification letters were issued to FreeStyle Navigator customers via certified mail and notification was posted on the Abbott Diabetes Care's Website. Customers were notified of the potential issue and actions to take: 1) if moisture is observed in the FreeStyle Navigator transmitter now or at any time, discontinue use of the transmitter and call Abbott Diabetes Care Customer Care at 1-800-418-6427 and 2) if a loss of connection is encountered or if your results are not reliable or are inconsistent with how you feel, consult your User's Guide for troubleshooting guidelines. If the issue is not resolved, call Abbott Diabetes Care Customer Care at 1-800-418-6421. For outside-of-US accounts, the firm's regional offices will distribute notices to identified customers in the appropriate language.
Worldwide Distribution -- United States (OH, FL, NJ, CA, MS, MN, MA, and NY), France, Australia, the Netherlands, Sweden, the UK, Israel, Ireland and Germany.
5,449 of total units