FDA Recall Terminated

ADVIA Chemistry Lipase Reagent , Catalog Number B01-4840-01 REF: 01984894.

Recall: Z-1457-2015 · Initiated March 23, 2015

Recall

Recall Number
Z-1457-2015
Event Number
70835
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2432235
Product Code
CHI
Status
Terminated
Root Cause
Device Design
Initiated
March 23, 2015
Posted
April 20, 2015
Terminated
December 18, 2015
Address
511 Benedict Ave, Tarrytown, NY, 10591-5005

Description

ADVIA Chemistry Lipase Reagent , Catalog Number B01-4840-01 REF: 01984894.

Reason

Siemens internal investigation confirmed that current contamination avoidance settings do not prevent carryover resulting in falsely elevated ADVIA Chemistry Lipase outlier results when running with Triglycerides, Triglycerides_2, Triglycerides_c, Cholesterol concentrated, and DLDL reagents.

Action

Siemens issued an Urgent Medical Device Correction on March 23, 2015, notifying direct consignees about the product, problem, and action to be taken. Customers were instructed to follow the directions provided in the customer letter in the event that a Lipase outlier is observed. All customers were requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Medical Device Correction letter within (30) days. Field service personnel have been sent an SKB (Service Knowledge Base) describing the issue and instructing them how to deal with customer questions. For further questions please call (312) 275-7795.

Distribution

Worldwide Distribution - US (nationwide) Distribution and to the countries of : Thailand, Pakistan, Vatican, Guadeloupe, Reunion, French Polynesia , Serbia, Martinique, India, China, Singapore, Brazil, South Africa, Mexico, Canada, Israel, Australia, Egypt, Russian Federation, Indonesia, Rep. Korea (S) ,Unit.Arab Emir., Taiwan ,Argentina ,Chile, Malaysia ,Colombia, Peru, Kazakhstan, Paraguay, Vietnam, Algeria, Austria, Belgium, Switzerland, Cyprus, Czech Republic, Germany, Estonia, Spain, Finland, France, United Kingdom, Greece, Hungary, Italy, Lithuania, Luxembourg, Latvia, Netherlands, Norway, Poland, Portugal, Romania, Slovenia, Slovakia and Turkey.

Quantity

7814