FDA Recall Terminated

Pioneer Surgical brand Quantum Spinal Rod System Locking Cap; Catalog No 10-LCAP.

Recall: Z-1457-05 · Initiated July 29, 2005

Recall

Recall Number
Z-1457-05
Event Number
32918
Firm
Pioneer Surgical Technology
FEI Number
1000115331
Product Code
MVH
Status
Terminated
Root Cause
Other
Initiated
July 29, 2005
Posted
August 31, 2005
Terminated
October 25, 2005
Address
375 River Park Cir, Marquette, MI, 49855-1781

Description

Pioneer Surgical brand Quantum Spinal Rod System Locking Cap; Catalog No 10-LCAP.

Reason

The product has a higher than normal locking torque, which may cause the saddle to malfunction or to break interoperatively.

Action

Distributors were notified via letter dated July 29, 2005 to retrieve all caps from inventory and point of use locations and to return them to Pioneer.

Distribution

Nationwide.

Quantity

5069