FDA Recall Terminated

Boston Scientific brand Concerto Bipolar Needle Electrode, sterile, latex free, Model M001262400, catalog number 26-240.

Recall: Z-1456-05 · Initiated July 15, 2005

Recall

Recall Number
Z-1456-05
Event Number
32764
Firm
Boston Scientific Corp
FEI Number
1828132
Product Code
GEI
Status
Terminated
Root Cause
Other
Initiated
July 15, 2005
Posted
August 31, 2005
Terminated
February 7, 2007
Address
780 Brookside Dr, Spencer, IN, 47460-1080

Description

Boston Scientific brand Concerto Bipolar Needle Electrode, sterile, latex free, Model M001262400, catalog number 26-240.

Reason

The distal tip may dislodge from the device during use and the firm has received complaints of electrical errors.

Action

Consignees were advised by recall letter dated July 25, 2005 and requested to return all units on hand.

Distribution

Nationwide, Germany, Italy, Netherlands, Spain and Sweden.

Quantity

115