Concerto & Basic Shower Trolly, Model Numbers BAB1000-01 (UDI: (01)05055982788083), BAB1000CON6102 [(01)05056097364483], BAB1007-01 [(01)05055982796255], BAB1101-01 [(01)05055982787949], BAB1102-01 [(01)05055982796156], BAB1103-01 [UDI: (01)05055982796163], BAB1107-01 [(01)05055982796187], BAB2000-01 [UDI: (01)05055982788106], BAB2101-01 [(01)05055982788007], BAB2102-01 [01)05055982796194], BAB3000-01 [(01)05055982788120], BAB3101-01 [(01)05055982788021], BAB3107-01[ (01)05055982796248], BAB5000-01 [UDI: (01)05055982788144]. and Concerto & Basic Shower Trolley Spare parts (safety catch) part no. 8451622, 8546608, S8533571-031, S8533572-031, S8533570-031 S8542201-014, S8523559-031,

Recall: Z-1452-2020 · Initiated August 22, 2019

Recall

Recall Number
Z-1452-2020
Event Number
84881
Firm
ARJOHUNTLEIGH POLSKA Sp. z.o.o. Ui. Ks. Piotra Wawrzyniaka 2 Komorniki Poland
FEI Number
3007420694
Product Code
FSA
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
August 22, 2019
Terminated
February 15, 2022

Description

Concerto & Basic Shower Trolly, Model Numbers BAB1000-01 (UDI: (01)05055982788083), BAB1000CON6102 [(01)05056097364483], BAB1007-01 [(01)05055982796255], BAB1101-01 [(01)05055982787949], BAB1102-01 [(01)05055982796156], BAB1103-01 [UDI: (01)05055982796163], BAB1107-01 [(01)05055982796187], BAB2000-01 [UDI: (01)05055982788106], BAB2101-01 [(01)05055982788007], BAB2102-01 [01)05055982796194], BAB3000-01 [(01)05055982788120], BAB3101-01 [(01)05055982788021], BAB3107-01[ (01)05055982796248], BAB5000-01 [UDI: (01)05055982788144]. and Concerto & Basic Shower Trolley Spare parts (safety catch) part no. 8451622, 8546608, S8533571-031, S8533572-031, S8533570-031 S8542201-014, S8523559-031,

Reason

The firm has become aware that any of the 4 safety catches may exhibit a mechanical failure (break in one of the arms of safety catch), which in certain circumstances may contribute to the side support unintentionally opening and may lead to a potential risk of a patient fall. The patient falling could lead to various types of injuries.

Action

On 08/22/2019 a "Urgent - Field Safety Notice Medical Device correction" letter was sent to all affected consignees for the Concerto & Basic Shower Trolley. The notification informed consignees about the recall and asked they take the following action: 1. Verify the condition of all 4 Safety Catches attached to your device. If you notice the break withdraw the trolley from use. 2. IMPORTANT: The device may stay in use until the upgrade is performed, providing that the following instructions are carefully followed: - It is recommended that the Safety Catches on the shower trolley Side Supports be inspected daily to ensure any of the catches are not broken as show on the Image - The patient is not left unattended at any time. - The patient is fully assisted so that he/she does not grab, push or lean against the Side Supports during showering activities. - Caregivers should limit pulling and pushing of the Side Supports with excessive force. - Note: always adhere to the Safe Working Load as outlined in the IFU. 3. Ensure that all caregivers and users of Arjo Concerto & Basic shower trolleys are made aware of this Field Safety Notice (FSN) and all listed devices at your facility are available to be retrofitted free of charge during the service technician visit. 4. If you have purchased spare parts listed, any of the following will be done free of charge: a. If the spare parts were sold to a 3rd party, please provide the sold to information on the Customer Response Form. Arjo will contact the 3rd party to make arrangements for repair or return of the affected parts. b. If the purchased parts were installed on a Concerto/Basic unit in your inventory, please provide the serial number on the Customer Response Form so an Arjo technician can inspect and retrofit the unit. c. If the purchased parts are still in your inventory, please complete and return the Customer Response Form and Arjo will contact you and make arrangements to replace the spare

Distribution

US: CA, CO, CT, GA, MA, MD, MI, NC, NJ, NY, OH, OK, TX, VA, WI OUS: Australia Austria, Bangladesh, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Iceland, India, Ireland, Italy, Japan, Korea, Luxembourg, Netherlands, Netherlands/CW, New Zealand, Norway, Philippines, Poland, Romaina, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Province of China, United Arab Emirates, United Kingdom

Quantity

669 trollys