FDA Recall Terminated

Biomet Ascent SI knee instruments sliding femoral 4-in-1 block, large, stainless steel; Ref. RD140632.

Recall: Z-1452-05 · Initiated May 16, 2005

Recall

Recall Number
Z-1452-05
Event Number
32124
Firm
Biomet, Inc.
FEI Number
1825034
Product Code
HTZ
Status
Terminated
Root Cause
Other
Initiated
May 16, 2005
Posted
August 31, 2005
Terminated
September 8, 2005
Address
56 E Bell Dr, Warsaw, IN, 46582-6989

Description

Biomet Ascent SI knee instruments sliding femoral 4-in-1 block, large, stainless steel; Ref. RD140632.

Reason

The instrument may have been assembled improperly, which may result in reversed resection cuts on the femur and affect implant performance, resulting in the need for revision surgery.

Action

Biomet distributors were issued a letter dated 5/16/05 instructing them to locate the instruments and conduct a field inspection of the instruments to verify the pin location offset and ensure the correct orientation of the anterior marking on the slide component.

Distribution

Nationwide, Australia, Austria, Belgium, Canada, Costa Rica, Denmark, Finland, Greece, Israel, Japan, Mexico, Netherlands, New Zealand, South Korea, Spain, and the United Kingdom.

Quantity

157