FDA Recall Terminated

Philips BuckyDiagnost Radiographic. Catalog Numbers: 704031, 704032, 704035, 704060, 704062 This system is used for making x-ray exposures for diagnostics

Recall: Z-1450-2013 · Initiated January 16, 2013

Recall

Recall Number
Z-1450-2013
Event Number
64128
Firm
Philips Healthcare Inc.
FEI Number
1218950
Product Code
KPR
Status
Terminated
Root Cause
Release of Material/Component prior to receiving test results
Initiated
January 16, 2013
Posted
June 17, 2013
Terminated
October 15, 2013
Address
3000 Minuteman Road, Andover, MA, 01810

Description

Philips BuckyDiagnost Radiographic. Catalog Numbers: 704031, 704032, 704035, 704060, 704062 This system is used for making x-ray exposures for diagnostics

Reason

An inspection of IATD (Installation Acceptance Test Document)IA SID (Source Image Distance) records shows table height measurements out of specification with the consequence of potential system mis-alignment

Action

Philips Healthcare issued an Urgent Field Safety Notification on January 11, 2013, to advised users of issued and perform corrective action to the units. Field Service Engineer (FSE) will execute corrections and attach record to IATD file. Your proposed corrective action plan (CAP) for the Bucky Diagnost system involves two mandatory field actions including new testing for sites with missing documentation and correcting the table height measurements for systems where the incorrect measurement procedure was used. You have also revised your IATD document to improve instructions to the field engineer for table height measurements. In your March 28, 2013, letter, you proposed a 6 month timeframe for this corrective action, and in your May 9, 2013 email, you communicated that the field corrections were ongoing and would be completed by July 16, 2013. In your May 9, 2013 email, you also clarified that the recalls are to be completed free of charge to any affected customer, whether or not they have a service agreement with Philips Healthcare. Further questions please call (978) 687-1501.

Distribution

Worldwide Distribution - US Nationwide and the country of Canada.

Quantity

116 units