FDA Recall
Terminated
Catalog Number: 9002469 BASIC SURGICAL PACK IV Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc.
Recall: Z-1448-2016
·
Initiated January 7, 2016
Recall
- Recall Number
- Z-1448-2016
- Event Number
- 72992
- Firm
- Customed, Inc Calle Igualdad Final # 7
- FEI Number
- 2648727
- Product Code
- LRO
- Status
- Terminated
- Root Cause
- Packaging
- Initiated
- January 7, 2016
- Posted
- April 14, 2016
- Terminated
- May 3, 2017
- Address
- Fajardo, PR, 00738
Description
Catalog Number: 9002469 BASIC SURGICAL PACK IV Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc.
Reason
Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.
Action
Recall Notification Letter sent to direct accounts via certified mail with return receipt on January 7, 2016 and January 8, 2016.
Distribution
Distributed only in Puerto Rico.
Quantity
30