FDA Recall Terminated

ADVIA 1650 Chemistry System, Automated Clinical Chemistry Analyzer - human serum, plasma and urine tests.

Recall: Z-1448-06 · Initiated February 17, 2006

Recall

Recall Number
Z-1448-06
Event Number
35880
Firm
Bayer Healthcare, LLC (Diagnostics Division)
FEI Number
2432235
Product Code
JGS
Status
Terminated
Root Cause
Other
Initiated
February 17, 2006
Posted
August 29, 2006
Terminated
May 24, 2012
Address
511 Benedict Avenue, Tarrytown, NY, 10591-5097

Description

ADVIA 1650 Chemistry System, Automated Clinical Chemistry Analyzer - human serum, plasma and urine tests.

Reason

Low QC recovery observed on carbamazepine (CARB) results immediately following a Gentamicin (GENT) assay or a Digoxin (DIG) assay. When GENT or DIG precedes CARB, the CARB test result is artificially low (as much as -25%). PHNY was determined to also exhibit a similar carryover effect on CARB. All others were found not to produce the same reagent probe-based carryover effect.

Action

Bayer HealthCare LLC issued a Support Bulletin to all affected Bayer HealthCare LLC Branches on/about 2/17/06 in order to inform them of the issue and to provide them with instructions for this ''field correction''. Customer Bulletins were also issued to the Branches both in the United States and elsewhere for communication with the affected customers.

Distribution

Worldwide including USA and Puerto Rico.

Quantity

1138 for all products