FDA Recall Terminated

Sybaritic, NannoLight MP50. Model Numbers and Description: 1.) 304207-MP 220 Volt, Multi-Platform configuration (able to attach optional laser handpiece when purchased separately), Subassembly (unit only, no safety glasses, owner's manual, etc. included); 2.) 404207 Completed unit (safety glasses, owner's manual, etc. included), glossy finish, 220 Volt; 3.) 404207-MP Completed unit, matte finish, 220 Volt, Multi-Platform configuration; 4.) 404207-B Completed unit, matte finish, 220 Volt; 5.) 404207-USAB Completed unit, matte finish, 220 Volt, USA-style power cord; 6.) 404208-B Completed unit, matte finish, 110 Volt; and 7.) 404208-MP Completed unit, matte finish, 110 Volt, Multi-Platform configuration. The NannoLight Intense Pulsed Light System is indicated for use in surgical, aesthetic and cosmetic applications requiring selective photothermolysis (photocoagulation or coagulation) and hemostasis of soft tissue in the medical specialties of general and plastic surgery and dermatology.

Recall: Z-1444-2009 · Initiated April 12, 2008

Recall

Recall Number
Z-1444-2009
Event Number
51704
Firm
Sybaritic, Inc
FEI Number
2131125
Product Code
GEX
Status
Terminated
Root Cause
Device Design
Initiated
April 12, 2008
Posted
September 3, 2009
Terminated
July 5, 2012
Address
9220 James Ave S, Bloomington, MN, 55431-2302

Description

Sybaritic, NannoLight MP50. Model Numbers and Description: 1.) 304207-MP 220 Volt, Multi-Platform configuration (able to attach optional laser handpiece when purchased separately), Subassembly (unit only, no safety glasses, owner's manual, etc. included); 2.) 404207 Completed unit (safety glasses, owner's manual, etc. included), glossy finish, 220 Volt; 3.) 404207-MP Completed unit, matte finish, 220 Volt, Multi-Platform configuration; 4.) 404207-B Completed unit, matte finish, 220 Volt; 5.) 404207-USAB Completed unit, matte finish, 220 Volt, USA-style power cord; 6.) 404208-B Completed unit, matte finish, 110 Volt; and 7.) 404208-MP Completed unit, matte finish, 110 Volt, Multi-Platform configuration. The NannoLight Intense Pulsed Light System is indicated for use in surgical, aesthetic and cosmetic applications requiring selective photothermolysis (photocoagulation or coagulation) and hemostasis of soft tissue in the medical specialties of general and plastic surgery and dermatology.

Reason

Sybaritic Inc. had an issue with the 585nm filter used with the NannoLight MP50. The filter may cause unwanted burning and should not be used for vascular lesion treatments.

Action

A "Sybaritic" notification letter dated April 15, 2008 was issued to consignees. The letter described the modifications to the protocols for the NannoLight MP50 and the temporary replacement of 585 Filter for the 560 Filter for Vascular Lesions until further notice. They requested the immediate use of the protocols replacing all other versions. A follow-up letter dated June 3, 2008 was issued to customers. The letter included an attachment " Product Notice Nannolight MP50 585 nm "special" filter from Sybaritic, Inc. The letter described the new 585 nm "bandwidth filter". Also the notification advised consignees to discontinue the use of the old 585 filter and to use 560nm filter until they receive the new 585 nm bandwidth filter. Please direct any questions to your Account Manager or to Sybaritic, Inc. by calling 1-800-445-8418 or 1-952-888-8282.

Distribution

Worldwide Distribution -- US (states of WA, CA, MN, FL and AR), POLAND, THAILAND, CHILE, CANADA, KOREA, AZERBAIJAN, INDIA, SPAIN, ARGENTINA, RUSSIA, GREECE, PHILIPPINES, POLAND, INDIA, JORDAN, TURKEY, MEXICO, URUGUAY,BELGIUM, LEBANON, BULGARIA and SINGAPORE.

Quantity

59