FDA Recall Terminated

Canister: CO2 canister with GE part numbers 1407-3200-000, 1407-7004-000, or M1084850. Intended to provide general inhalation anesthesia and ventilatory support.

Recall: Z-1443-2013 · Initiated May 14, 2013

Recall

Recall Number
Z-1443-2013
Event Number
65164
Firm
GE Healthcare, LLC
FEI Number
2126677
Product Code
BSZ
Status
Terminated
Root Cause
Component change control
Initiated
May 14, 2013
Posted
June 3, 2013
Terminated
March 12, 2014
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

Canister: CO2 canister with GE part numbers 1407-3200-000, 1407-7004-000, or M1084850. Intended to provide general inhalation anesthesia and ventilatory support.

Reason

GE Healthcare has recently become aware of a potential safety issue involving the reusable CO2 absorbent canister accessory used with Aespire, Aespire View, Avance, Avance CS2, Aisys, ADU, 9100, M900, M904, 9300 systems. Reusable CO2 absorbent canisters accessories may not seal properly.

Action

GE Healthcare sent an "URGENT MEDICAL DEVICE CORRECTION" letter April 30, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. A Fax back form was included with the letter requesting the consignees to complete and return.

Distribution

Worldwide Distribution-USA (nationwide) including the states of AZ, CA, CO, CT, ID, MK, MI, NJ, NY, OH, TN, TX, and VA, and the countries of INDIA, AUSTRALIA, BAHRAIN, BANGLADESH, BOLIVIA, BOSNIA & HERZEGOWINA, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, HONG KONG, HUNGARY, INDONESIA, IRAQ,ISRAEL, ITALY, JAPAN, REPUBLIC OF KOREA,LEBANON, MALAYSIA, MEXICO, NETHERLAND,NEW ZEALAND, OMAN, PARAGUAY, PERU., PHILIPPINES, RUSSIAN FEDERATION, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, THAILAND, TURKEY, UNITED ARAB EMIRATES , UNITED KINGDOM, and VIET NAM.

Quantity

800