FDA Recall Terminated

ProTrack Microcatheter; Product Usage: The ProTrack Microcatheter is a sterile, single use device intended to be used as a standard infusion wire or as an exchange guide for exchanging guidewires.

Recall: Z-1442-2015 · Initiated March 30, 2015

Recall

Recall Number
Z-1442-2015
Event Number
70896
Firm
Baylis Medical Corp 5959 TransCanada Highway Montreal Canada
FEI Number
3005241199
Product Code
DQX
Status
Terminated
Root Cause
Process control
Initiated
March 30, 2015
Posted
April 13, 2015
Terminated
November 17, 2016

Description

ProTrack Microcatheter; Product Usage: The ProTrack Microcatheter is a sterile, single use device intended to be used as a standard infusion wire or as an exchange guide for exchanging guidewires.

Reason

Baylis Medical Company is recalling two lots of ProTrack Microcatheter due to the possibility that the catheter may have circumferential defects along its shaft.

Action

A recall notification, marked "URGENT" will be sent by Baylis Medical via overnight courier to all customers affected by the recall with a fax-back form. The form is to be faxed back to Baylis Medical confirming receipt of the recall notice. As per instructions on the fax-back form, affected devices are requested to be returned. Baylis Medical will replace the returned affacted catheters with a new catheter.

Distribution

US Nationwide Distribution - US Nationwide in the states of MO, MA, OH, CA, AR, MA, FL, GA, CO, TX, KY, NY, MI, WI, OR

Quantity

58 units